Determination of Iron in Biological Materials by inductively coupled plasma mass spectrometry (ICP-MS)

Authors

  • A. Kaňa Department of Analytical Chemistry, Institute of Chemical Technology, Prague
  • O. Mestek Department of Analytical Chemistry, Institute of Chemical Technology, Prague

Keywords:

iron, ICP-MS, DRC

Abstract

Determination of iron by the title method is complicated by spectral interferences of polyatomic ions, in particular ArO+ and CaO+. This work was focused on the possibility of reducing the interference using a dynamic reaction cell (DRC). The cell was pressurized with He (collision gas) and methane (reaction gas). Settings of DRC were optimized to achieve maximal efficiency of elimination of interferences. The determination of 56Fe and 57Fe was tested and the limit of detection and linearity of calibration curves were determined. The accuracy of the method was verified by analysis of a certified reference material.

Published

2012-03-15

How to Cite

Kaňa, A., & Mestek, O. (2012). Determination of Iron in Biological Materials by inductively coupled plasma mass spectrometry (ICP-MS). Chemické Listy, 106(3), 211–216. Retrieved from http://www-.chemicke-listy.cz/ojs3/index.php/chemicke-listy/article/view/967

Issue

Section

Articles

Most read articles by the same author(s)

1 2 > >>